Designed for the analysis of anions and cations
The IONUS system is suitable for the analysis of anions and cations in research, development, and routine applications.
Sample solutions are primarily aqueous-based and may contain water-miscible organic solvents. A classic example is the analysis of (drinking) water in accordance with EPA Method 300, Parts A and B.
The advantage of IONUS is its compatibility with IC columns from all manufacturers and with all common eluents used in anion and cation chromatography.
Key technical specifications of the system
METHOD
Ion exchange chromatography with an integrated conductivity detector; optional UV detection available
Conductivity detection
Noise: 0.02 µS/1 s | Range: 0 – 10 mS/cm
Reproducibility
0.1% RSD at 100 µL/min
Flow scheme
Suppressor technology can be added to the instrument to measure lower concentration ranges during anion analysis.
In applications performed without suppressor technology, the chromatogram may exhibit both negative and positive peaks for each substance, or peak areas may be very small due to minimal conductivity differences.
In such cases, an external UV/VIS detector can provide a solution by employing indirect UV detection for light-absorbing eluents.
In such cases, an external UV/VIS detector can provide a solution by employing indirect UV detection for light-absorbing eluents.
Another possibility is the use of non-absorbing eluents for light-absorbing sample substances.

Technical Details

MethodIon-exchange chromatography of anions and cations
Manual injection20 µL; additional injection loops (50 µL, 100 µL, 200 µL) available upon request
Autosampler2 racks (each accommodating 48 vials [1.5 mL] or 12 vials [6 mL]), sample cooling, 100 µL loop, 3 injection modes (full loop, partial loop, microliter pickup)
ColumnStainless steel / PEEK
Temperature stability / range~ 0.5 °C / ambient temperature up to 50 °C
DetectionConductivity detector: noise: 0.02 µS, 1s | linearity: < 1 s | range: 0 - 10 mS/cm | effective cell volume: 1 µL, conductivity block: 30 °C
Isocratic pump systemMax. pressure: 400 bar | flow rate: 0.01 - 10.00 mL/min | flow rate reproducibility: 0.1% RSD at 100 µL/min
Maximum operating pressure120 bar
Control software and data acquisitionClarity (21 CFR Part 11 compliant) or AminoPeak
Dimensions, weight, power1000 x 350 x 375 mm*, 60 kg, 110 - 230 V, 50/60 Hz