Freely definable formulas
Applicable standardsISO 4406 (91/99), NAS 1638, GB 5930-86; GJB-420A-96, GJB-420B, GOST 17216-71, SAE-A6D, SAE-749D, SAE-AS4059E, or 16 freely definable size channels (up to 3 formulas selectable on the device; downloadable and configurable via "Protrend" evaluation software)
Measured value outputPC, display, printer, threshold output; data memory for 600 measurements
SoftwareWhen using "Protrend" evaluation software on a PC, various methods with up to 32 size classes can be applied. Presentation of measurements in tables and diagrams is possible (with Excel export)
Data memory for 600 measurements
DPS pumpDisplacement volume: 10 ml; Flush/measurement volume: 10–1000 ml; Inlet pressure: 10 bar; Option: Pump operation in suction or pressure mode up to 315 bar via DMV adapter block
Ambient temperature5–40°C / Fluid temperature: 0–70°C
Power supply12 VDC/1 A Friwo plug-in power supply
ConnectionMinimess (M 16 x 3); 6 mm Ermeto (L)
Dimensions of PCSS Fluid Lite (L x W x H)280 x 130 x 270 mm
AccessoriesSample feeder for pressurizing sample bottles